Calibrating a scanning electron microscope (SEM)

Introduction
Although modern manufacturing techniques have greatly improved the accuracy of SEMs, it is good practice to regularly check if the magnification shown on the SEM is still correct.  One should also take into account that magnification in the SEM changes with working distance. Magnification and the size marker shown on the SEM image are a best effort, but not always correct.  The magnification error of an SEM can be in the range of 5-10%. Calibration becomes of utmost importance if measurements are made on SEM images.


SEM calibration standards
The correct calibration feature (size) needs to be used for calibrating at a certain magnification.  Available EM-Tec SEM calibration standards are:
EM-Tec LAMS-15 for low magnification calibrations (15mm down to 10µm)
EM-Tec M1 and M10 with 1 and 10µm patterns for medium range magnification calibrations
EM-Tec MCS-1 for medium magnification range calibrations  (2.5mm down to 1µm)
EM-Tec MCS-0.1 for medium to high magnification calibrations (2.5mm down to 100nm)

 

Instructions on calibration
The magnification calibration standards comprise features, mostly lines with a known distance (pitch). It is best to select a calibration feature size compatible with the magnification used for imaging. A calibration for 1000x magnification is not valid for 100x or 10,000x magnification. When imaging is in the 2000x – 3000x magnification, calibrate at 2500x maginfication.

Calibration procedure steps:

  1. Select the calibration distance on the calibration standard – for example 10µm and make sure that the lines are vertical on the screen.
  2. Select condensor lens setting and keep this setting throughout the calibration and imaging procedure
  3. Always remove lens hysteresis or approach always either from low or from high magnification
  4. Image the 10µm calibration feature to cover at least 10-20% in the middle of the image using an easy number like 2000 or 5000x magnification
  5. Measure the pitch of the lines horizontally (either center to center or left to left or right to right). Ideally this should be 10um if the system is calibrate correctly. In service or administrator mode, most SEMs offer the option to correct either magnification or size marker distance.

Correct magnification or measured size
To correct magnification or measured size multiply use the following formulas:
Magnification correction:      Mcalibitrated = Mshown x Dshown/Dcalibrated
Measurement correction:     Dactual = Dmeasured x Dcalibrated/Dshown

Where:
Mcalibrated  = Calibrated magnification
Mshown = Magnification shown on SEM
Dcalibrated  = Actual feature size of calibration standard
Dshown  = Feature size of calibration standard shown on SEM
Dactual  = Actual (calibrated) distance in SEM image
Dmeasured   = Distance measured in SEM image

 

Example for magnification calibration:
Magnification of 5000x (Mshown) shows the 10µm distance(Dcalibrated) as 11µm (Dshown)
The calibrated magnification (Mcalibrated) in reality is then:  5000x11/10 = 5500x. The magnification correction factor is 1.1


Example calibrated features sizes in images:
A feature in the same image as above is 25µm (Dmeasured). Previously, it has been determined that the 10µm on the calibration standard (Dcalibrated) was shown as 11µm (Dshown) on the SEM image (screen).
The correct feature distance (Dactual) would be 25 x10/11 = 22.73µm. The correction factor for size measurements is this case is 0.91.


Notes for SEM calibrations:

  • Only calibrate with a stable SEM; system and beam correctly saturated on for at least 30-60 minutes
  • Calibrate at a specific working distance and SEM settings and keep these for subsequent measures. When changing working distances and/or settings, calibrate again.
  • Always approach focus and condenser lens settings from the same side to counter hysteresis in the lenses; always from high to low or the other way around.
  • The calibration standard contains a certificate; in the certificate you will find the calibrated feature sizes of the calibration features and the uncertainty (precision). For example, the nominal 10um feature might be calibrated at 10.05µm; use the 10.05µm for calibration purposes.
  • The calibration uncertainty of an SEM is always equal or of higher uncertainty than that of the calibration standard.
  • Check against image or stage drift; do not attempt calibration when drift is present.
  • Provide a sufficient signal to produce clear images for calibration measurements
  • Consider having the SEM calibrated by a qualified service engineer during maintenance



SEM Supplies

TEM Supplies

Calibration

Sample Preparation

AFM / SPM

  Product Menu with Images
  SEM Sample Stubs
  Carbon Tabs
  Sample Stub Adapters
  SEM Stage Adapters
  SEM Sample Holders
  SEM Preparation Stands
  Filaments / Cathodes
  Silicon Finder Grid
  Gatan 3View Pins
  FEI Volumescope Pins
  SEM Stub Storage Boxes
  Phenom Supplies
  JEOL NeoScope Supplies
  Hitachi TM Series Supplies
  Field & Lab Sampler Kits
  FlowView liquid sample Kit

Instruments

  Acoustic Enclosures
  Cressington SEM coaters
  TEM Sample Prep
  Diaphragm Pump
  Rotary vacuum Pump
  Precision diamond saw  
  TEM Grids
  TEM Support Films
  Silicon Nitride Films
  K-Kit Wet Cell Holder
  Graphene Films
  TEM Grid Boxes
  Cryo Grid Boxes
  TEM Sample Staining
  3mm Embedding Tubes
  Filaments/Cathodes
  Eyelash Manipulators

Cryo Supplies

  Cryo Grid Boxes
  Other Cryo Supplies

FIB Supplies

  FIB Lift-out Grids
  FIB Low Profile Stubs
  FIB Grid Holders
  FIB Pre-tilt Holders
  FIB Pre-tilt stubs
  FIB Grid Boxes
  SEM / FIB Magn. Calib.
  SEM Res. Test Spec.
  EDS / WDS Calibration
  TEM Calibration
  AFM / SPM Calibration
  LM Magn. Calibration

Vacuum Supplies

  Vacuum Gauges
  KF/NW Vacuum Parts
  KF/NW Vacuum Hoses
  ISO Vacuum flange parts
  Vacuum Oil and Grease
  Vacuum Sealant
  Diaphragm Pump
  Rotary vacuum Pump
  Vacuum Sample Storage
  Vacuum Pick-Up Pens

Sample Coating Supplies

  Sputter Targets
  Carbon Rods & Fibers
  Quartz QCM Crystals
  SEM Sample Coating Fluid

Evaporation Supplies

  Thermal Evaporation Sources
  Evaporation Materials
  Supports & Substrates
  SEM Preparation Stands
  Tweezers
  Probes & Picks
  Applicators & Swabs
  Plastic Transfer Pipettes
  Cutting Tools / Scissors
  Conductive Paint/Cement
  Conductive Tapes & Tabs
  Non-Conductive Adhesives
  Sample Storage
  Hand Tools
  Sorting Wire Mesh
  Preparation Surfaces
  Cleaning / Gloves
  Conductive Metal Powders
  Metallographic Sample Prep
  AFM/SPM Discs
  AFM/SPM Disc Pick-up Tool
  AFM/SPM Holders
  AFM/SPM Disc Storage
  Cantilever Tweezers
  AFM/SPM Discs Tweezers
  Nano-Tec Mica Discs
  HOPG Substrates

Light Microscopy

  Correlative Cover Slips
  Glass Microscope Slides
  Glass Cover Slips
  Quartz Microscope Slides
  Quartz Cover Slips
  Black Metal Slides
  Slide Storage Boxes
  LM Calibration
  Plastic Transfer Pipettes
  Optical Lens Tissue
  Glass Petri Dishes
  Digital Microscope
  PTFE Beakers