Introduction
Although modern manufacturing techniques have greatly improved the accuracy of SEMs, it is good practice to regularly check if the magnification shown on the SEM is still correct. One should also take into account that magnification in the SEM changes with working distance. Magnification and the size marker shown on the SEM image are a best effort, but not always correct. The magnification error of an SEM can be in the range of 5-10%. Calibration becomes of utmost importance if measurements are made on SEM images.
SEM calibration standards
The correct calibration feature (size) needs to be used for calibrating at a certain magnification. Available EM-Tec SEM calibration standards are:
EM-Tec LAMS-15 for low magnification calibrations (15mm down to 10µm)
EM-Tec M1 and M10 with 1 and 10µm patterns for medium range magnification calibrations
EM-Tec MCS-1 for medium magnification range calibrations (2.5mm down to 1µm)
EM-Tec MCS-0.1 for medium to high magnification calibrations (2.5mm down to 100nm)
Instructions on calibration
The magnification calibration standards comprise features, mostly lines with a known distance (pitch). It is best to select a calibration feature size compatible with the magnification used for imaging. A calibration for 1000x magnification is not valid for 100x or 10,000x magnification. When imaging is in the 2000x – 3000x magnification, calibrate at 2500x maginfication.
Calibration procedure steps:
Correct magnification or measured size
To correct magnification or measured size multiply use the following formulas:
Magnification correction: Mcalibitrated = Mshown x Dshown/Dcalibrated
Measurement correction: Dactual = Dmeasured x Dcalibrated/Dshown
Where:
Mcalibrated = Calibrated magnification
Mshown = Magnification shown on SEM
Dcalibrated = Actual feature size of calibration standard
Dshown = Feature size of calibration standard shown on SEM
Dactual = Actual (calibrated) distance in SEM image
Dmeasured = Distance measured in SEM image
Example for magnification calibration:
Magnification of 5000x (Mshown) shows the 10µm distance(Dcalibrated) as 11µm (Dshown)
The calibrated magnification (Mcalibrated) in reality is then: 5000x11/10 = 5500x. The magnification correction factor is 1.1
Example calibrated features sizes in images:
A feature in the same image as above is 25µm (Dmeasured). Previously, it has been determined that the 10µm on the calibration standard (Dcalibrated) was shown as 11µm (Dshown) on the SEM image (screen).
The correct feature distance (Dactual) would be 25 x10/11 = 22.73µm. The correction factor for size measurements is this case is 0.91.
Notes for SEM calibrations: